Ticket #712 (closed defect: fixed)
Opened 2012-06-11T16:00:14-05:00
Last modified 2012-06-15T18:44:29-05:00
Tri-exponential fit results don't agree with TRI2
Reported by: | aivar | Owned by: | aivar |
---|---|---|---|
Priority: | major | Milestone: |
|
Component: | slim-plugin | Severity: | serious |
Keywords: | Cc: | ||
Blocked By: | Blocking: | #674 |
Description (last modified by aivar)
Using same RLD+LMA fit as #697, (146, 230) of 100611-YG-256.sdt, Gaussian noise:
Single exponential results agree, but triple exponential differ:
TRI2: A1 73.24 T1 2.09 A2 3.87 T2 1.77 A3 2.82 T3 0.00E+0 X2 0.852904
SLIM Plugin: A1 87.63362 T1 1.911811 A2 30.0 T2 0.25 A3 30.0 T3 0.1 Z 1.1207124 X2 0.8863228
Interesting that SP A2/T2 and A3/T3 results are so bad. They look like initial estimates that did not change.